An Intact ACTH LC-MS/MS Assay as an Arbiter of Clinically Discordant Immunoassay Results.

Quantifying biologically active ACTH


BACKGROUND: Measurement of plasma adrenocorticotropic hormone (ACTH) is key in the differential diagnosis of hypothalamic-pituitary-adrenal disorders. Two-site sandwich immunoassays dominate clinical testing of ACTH in North America; however, discordant results between manufacturers have been repeatedly reported. To resolve the discrepancy, we developed a liquid chromatography–tandem mass spectrometry (LC-MS/MS) assay for the intended measurand, biologically active intact ACTH (iACTH).


METHODS: The multiple reaction monitoring LC-MS/MS assay was designed to selectively measure full-length iACTH, as well as ACTH analogs and fragments (i.e., ACTH1–24 and ACTH18–39). Epitope assignment of the Roche Elecsys antibodies was performed by MALDI-TOF mass spectrometry. A method comparison between Roche Elecsys and Siemens Immulite ACTH immunoassays was performed and clinically concordant/discordant results identified. In a subset of these samples, the iACTH concentration was determined using the LC-MS/MS method.


RESULTS: The lower limit of the measuring interval of the iACTH LC-MS/MS assay was 9 pg/mL (2 pmol/L). The assay was linear from 9 to 1938 pg/mL (2 to 427 pmol/L). Epitope mapping revealed that the Roche capture and detection antibodies bound residues 9–12 and 36–39 of ACTH, respectively. The iACTH LC-MS/MS analysis demonstrated that for discordant results between 2 immunoassays studied, only the Roche results were highly positively correlated with the iACTH concentration.


CONCLUSION

Immunoprecipitation of biologically active ACTH molecules followed by LC-MS/MS analysis enabled selective detection of iACTH and relevant biologically active fragments in plasma. Applied to the investigation of clinically discrepant results, this method can act as an arbiter of the concentration of iACTH present.



bottom of page